COA Checklist and Lot-Matching Practice

Use the tool in your browser. The page processes your entries locally and does not save them.

Certificate reading checklist

Compare the lot on the container label with the lot on the certificate. Mark the report fields you find and save a text copy of your review.

Compare lot identifiers

Enter up to 180 characters per lot. The comparison ignores spaces at the start and end. Case, internal spacing, prefixes and leading zeros matter.

Both lot identifiers are missing.

Fields to locate

0 of 6 fields marked as located.

Practice with the fictional reports

Loading a case replaces the current review. Read the two reports and labels.

This review records lot text and document fields. Verifying a certificate and linking it to a received container require supporting records.

Read the reports and compare the labels

Fictional teaching reports. All issuers, identifiers, results and criteria below are simulated.

Certificate of analysis · TCOA-024

Report TCOA-024 header and sample record
FieldFictional entry
Report and authorizationTCOA-024 / Rev 01 / page 1 of 1. Issued by Training Lab 04, authorized by teaching reviewer TR-04.
MaterialLinear hexapeptide GAGKLV with free termini. Catalog EDU-HX06. Nominal 5.0 mg peptide equivalent per vial.
Tested sampleOne sealed vial submitted by Classroom Supplier 02. Lab accession EDULAB-0251 is mapped to supplier lot EDU-L-0042. Individual vial ID is unreported.
DatesReceived 2026-08-10, tested 2026-08-12, issued 2026-08-13.
Supporting recordsReferenced LC–MS spectrum TCOA-024-M, integration TCOA-024-H, and assay calculation TCOA-024-A. These attachments are outside the exercise.
TCOA-024 test results and stated acceptance criteria
Analyte or attributeMethod and basisUnitResultAcceptance criterion
GAGKLV mass matchLC–MS M-07 Rev 02, deconvoluted neutral monoisotopic massDa543.339543.338 ± 0.010 Da
Main chromatographic peakRP-HPLC/UV H-04 Rev 03 at 214 nm, 984 of 1,000 integrated signal-area unitsarea %98.4≥97.0 area %
GAGKLV amount in the tested vialCalibrated HPLC/UV assay A-02 Rev 01, assigned-content referencemg peptide equivalent per vial4.704.50–5.50 mg per vial
EndotoxinMethod unreportedUnreportedNot reportedUnreported

The mass differs from the stated target by +0.001 Da. The area percentage uses the integrated UV signal as its denominator. The reported amount comes from a separate calibrated assay. Ask the issuer about the scope of any unreported test.

Second report · TCOA-027

Training Lab 04 issued TCOA-027 / Rev 02 / page 1 of 1 on 2026-08-20 after testing on 2026-08-19. It lists material EDU-HX06 and lab accession EDULAB-0257. The supplier lot and the intake record that would link it to the accession are unreported.

TCOA-027 result for the submitted specimen
Analyte or attributeMethod and basisUnitResultAcceptance criterion
GAGKLV mass matchLC–MS M-07 Rev 02, deconvoluted neutral monoisotopic massDa543.340543.338 ± 0.010 Da
Three EDU-HX06 labels to compare
Vial labelLabel lotReport / accessionReported lotDocument finding
A · BOX2-05EDU-L-0042TCOA-024 Rev 01 / EDULAB-0251EDU-L-0042Exact lot match. The individual vial link remains open.
B · BOX2-06EDU-L-0044TCOA-027 Rev 02 / EDULAB-0257UnreportedMissing link. Ask for the accession-to-lot record.
C · BOX2-07EDU-L-0043TCOA-024 Rev 01 / EDULAB-0251EDU-L-0042Lot conflict. Request the correct report or a documented correction.

Practice the three comparisons and export your review.